NANOSENORS qp-fast AFM probes with its 3 cantilevers are designed for soft-, standard- and fast-non-contact or tapping mode AFM imaging (also known as attractive or dynamic mode). This AFM probe combines high operation stability with outstanding sensitivity and fast scanning ability in air and liquid environments.

A metallic layer (Au) is coated on the detector side of the cantilever. The virtually stress-free coating is bending the cantilever less than 2 % of the cantilever length.

The NANOSENSORS uniqprobe combines the well-known features of the other NANOSENSORS AFM probe series such as high application versatility and compatibility with most commercial SPMs with the additional advantage of a strongly reduced dispersion of force constant and resonance frequency.

The unsurpassed uniformity of the mechanical characteristics of the uniqprobe series is particularly important for applications, where a large number of probes with known and near identical force constants or resonance frequencies are needed.

The probe offers unique features:

Cantilever data:

Property Nominal Value Specified Range
Resonance Frequency /kHz
CB2: 420
CB3: 250
CB1: 800
CB2: 340 - 500
CB3: 200 - 300
CB1: 600 - 1000
Force Constant /(N/m)
CB2: 30
CB3: 15
CB1: 80
CB2: 20 - 45
CB3: 10 - 20
CB1: 50 - 140
Thickness /nm
CB1: 2500
CB2: 2500
CB3: 2500
CB1: 2470 - 2530
CB2: 2470 - 2530
CB3: 2470 - 2530
Mean Width /µm
CB1: 22
CB2: 27
CB3: 32
CB1: 20 - 24
CB2: 25 - 29
CB3: 30 - 34
Length /µm
CB1: 40
CB2: 60
CB3: 80
CB1: 35 - 45
CB2: 55 - 65
CB3: 75 - 85

Order codes and shipping units:

Order Code Number of AFM probes per pack Data sheet Coating
qp-fast-10 10 without partial Au coating
qp-fast-20 20 without partial Au coating
qp-fast-50 50 without partial Au coating


 

Special handling information for NANOSENSORS™ uniqprobes

Due to their unique geometry the tips of the uniqprobes are more susceptible to tip damage by electrostatic discharge (ESD) than other Silicon-SPM-Probes.

Electric fields near the probe chip may lead to field evaporation which can blunt the tip apex of the probe tip. Therefore the NANOSENSORS™ uniqprobes are shipped in specially designed ESD-safe chip carriers.

NANOSENSORS™ recommends to their customers to take appropriate precautions to avoid tip damage due to electrostatic discharge when handling the probes. This can for example be done by using anti-electrostatic mats, wrist bands and tweezers.

Product screencast NANOSENSORS™ uniqprobe AFM Probes series

Please watch the new Product Screencast on the NANOSENSORS™ uniqprobe AFM Probes series by product developer Dr. Laure Aeschimann at YouTube
Please subscribe to the NANOSENSORS Youtube Channel.


For further information please contact your local distributor or NANOSENSORS™ directly.
NANOSENSORS™ is a trademark of NanoWorld AG / info@nanosensors.com / www.nanosensors.com


NANOSENSORS™ is a trademark of NanoWorld AG - Copyright by NanoWorld AG